CSA INTERNATIONAL
Certificate of Com lian
p ce
Certificate: 1665151 (LR 105802) Master Contract: 162767
Project: 2372993 Date Issued: November 24,2010
Issued to: KROHNE,Inc.
7 Dearborn Rd
Peabody,MA 01960
USA
Attention: Rick Loiselle
The products listed below are eligible to bear the CSA Mark shown
ELLWoL4-d,Fo-o-
® Issued by: Edward Foo, C.E.T.
PRODUCTS
CLASS 2258 03 -PROCESS CONTROL EQUIPMENT-Intrinsically Safe and Non-
Incendive Systems-For Hazardous Locations
Class I,Division 2, Groups A,B,C and D; Class II,Division 2,Groups F and G:
• OPTIFLUX 1000-DIV2 series Primary Head.Model VN174abcdefghijklmn.
Operating ambient-50°C to 40°C; Temperature Code T6. Enclosure type as specified in the detail below.
Where:
a=Nominal Diameter: 5, 6, 8,B, C,E,F, or H.
b=Nominal Pressure: A, or B.
c=Approval: C.
d=System design: 1,2,4, 6,A,or C.
e=Converter Model: C, or D.
f=Mounting Material: (any alphanumeric digit).
g=Grounding Ring Material: (any alphanumeric digit).
DQD 507 Rev.2009-09-01 Page:1
SR
CSA INTERNATIONAL
Certificate: 1665151 (LR 105802) Master Contract: 162767
Project: 2372993 Date Issued: November 24,2010
h=Electrodes: 0.
i=Protection Class:
0 indicates painted aluminum enclosure,rated Type 4X and Type 6.
2 indicates stainless steel enclosure,rated Type 6P.
j =Cable: 2(Type 6P),or any other alphanumeric digit(Enclosure types other than 6P).
k=Cable Length: (any alphanumeric digit).
1=Calibration: (any alphanumeric digit).
in=Construction Requirements: (any alphanumeric digit).
n=QA/QC Requirements: (any alphanumeric digit).
Notes:
1) OPTIFLUX Primary Head must be used with the IFC 300 F-DIV2,IFC 300 F(/i)-DIV2, or IFC 090-F
DIV2 signal converters.
2) I.S. Electrodes when used with IFC 300 F-DIV2, IFC 300 F(/i)-DIV2, or IFC 090-F DIV2 signal
converters.
• OPTIFLUX 2000-DIV2 series Primary Head.Model VN144abcdefghijklmnop.
Operating ambient-50°C to 40°C; Temperature Code T6. Enclosure type as specified in the detail below.
Where:
a=Nominal Diameter: 4, 5, 6,7, 8,A,B,C, or D.
b=Nominal Pressure: 3,4, 5,A, or B.
c=Approval: C.
d=Cable Connection: 1,2,4, 6,A, or C.
e=Converter Model: C, or D.
DQD 507 Rev.2009-09-01 Page:2
SR
CSA INTERNATIONAL
Certificate: 1665151 (LR 105802) Master Contract: 162767
Project: 2372993 Date Issued: November 24,2010
f=Lining: (any alphanumeric digit).
g=Electrode: 1,3, or 6.
h=Construction of Electrodes: 1.
i=Housing/Flange material: 1, 3,or 4.
j =Protection Class/dimension:
0 indicates painted aluminum enclosure,rated Type 4X and Type 6.
2 indicates stainless steel enclosure,rated Type 6P.
k=Cable: 2 (Type 6P), or any other alphanumeric digit(Enclosure types other than 6P).
1=Cable Length: (any alphanumeric digit).
m=Calibration: (any alphanumeric digit).
n=G-P Ring/Material: (any alphanumeric digit).
o=Special: 0.
p=Module/labeled: 1
Notes:
1) OPTIFLUX Primary Head must be used with the IFC 300 F-DIV2,IFC 300 F(/i)-DIV2, or IFC 090-F
DIV2 signal converters.
2) I.S. Electrodes when used with IFC 300 F-DIV2, IFC 300 F(/i)-DIV2, or IFC 090-F DIV2 signal
converters.
• OPTIFLUX 2000-DIV2 series Primary Head. Model VN154abcdefghijklmnop.
Operating ambient-50°C to 40°C;Temperature Code T6. Enclosure type as specified in the detail below.
Where:
a=Nominal Diameter: E,F,G,H,K,L,M,or N.
b=Nominal Pressure: 2, 3,4, 5,A, or B.
c=Approval: C.
DQD 507 Rev.2009-09-01 Page:3
SP
CSA INTERNATIONAL
----------------------
Certificate: 1665151 (LR 105802) Master Contract: 162767
Project: 2372993 Date Issued: November 24,2010
d=Cable Connection: 1,2,4, 6,A, or C.
e=Converter Model: C, or D.
f=Lining: (any alphanumeric digit).
g=Electrode: 1, 3, or 6.
h=Construction of Electrodes: 1.
i=Housing/Flange material: 1, 3, or 4.
j =Protection Class/dimension:
0 indicates painted aluminum enclosure,rated Type 4X and Type 6.
2 indicates stainless steel enclosure,rated Type 6P.
k=Cable: 2 (Type 6P), or any other alphanumeric digit(Enclosure types other than 6P).
1=Cable Length: (any alphanumeric digit).
m=Calibration: (any alphanumeric digit).
n=G-P Ring/Material: (any alphanumeric digit).
o=Special: 0.
p=Module/labeled: 1
Notes:
1) OPTIFLUX Primary Head must be used with the IFC 300 F-DIV2, IFC 300 F(/i)-DIV2, or IFC 090-F
DIV2 signal converters.
2) I.S. Electrodes when used with IFC 300 F-DIV2, IFC 300 F(/i)-DIV2, or IFC 090-F DIV2 signal
converters.
• OPTIFLUX 2000-DIV2 series Primary Head. Model VN I 64abcdefghijklmnop.
Operating ambient-50°C to 40°C; Temperature Code T6. Enclosure type as specified in the detail below.
DQD 507 Rev.2009-09-01 Page:4
SR
CSA INTERNATIONAL
Certificate: 1665151 (LR 105802) Master Contract: 162767
Project: 2372993 Date Issued: November 24,2010
Where:
a=Nominal Diameter:Up to a maximum of 120 inches at 276 PSI.
b=Nominal Pressure: 1,2, or A.
c=Approval: C.
d=Cable Connection: 1,2,4, 6,A, or C.
e=Converter Model: C, or D.
f=Lining: (any alphanumeric digit).
g=Electrode: 1,3, or 6.
h=Construction of Electrodes: 1.
i=Housing/Flange material: 1, 3,or 4.
j =Protection Class/dimension:
0 indicates painted aluminum enclosure,rated Type 4X and Type 6.
2 indicates stainless steel enclosure,rated Type 6P.
k=Cable: 2 (Type 6P), or any other alphanumeric digit(Enclosure types other than 6P).
1=Cable Length: (any alphanumeric digit).
m=Calibration: (any alphanumeric digit).
n=G-P Ring/Material: (any alphanumeric digit).
o=Special: 0.
p=Module/labeled: 1
Notes:
1) OPTIFLUX Primary Head must be used with the IFC 300 F-DIV2, IFC 300 F(/i)-DIV2, or IFC 090-F
DIV2 signal converters.
2) I.S.Electrodes when used with IFC 300 F-DIV2,IFC 300 F(/i)-DIV2,or IFC 090-F DIV2 signal
converters.
DQD 507 Rev.2009-09-01 Page:5
SR
CSA INTERNATIONAL
Certificate: 1665151 (LR 105802) Master Contract: 162767
Project: 2372993 Date Issued: November 24,2010
• OPTIFLUX 4000-DIV2 series Primary Head. Model VN024abcdefghijklmnop.
Operating ambient-50°C to 40°C; Temperature Code T6. Enclosure type as specified in the detail below.
Where:
a=Nominal Diameter: 1,2, 3, 5, 6, or 7.
b=Nominal Pressure: A, or B.
c=Approval: C.
d=Cable Connection: 1,2,4, or A.
e=Converter Model: C, or D.
f=Lining: (any alphanumeric digit).
g=Electrode: 1,2, 3,4, 5, 6, 7,G,or U.
h=Construction of Electrodes: 1.
i=Housing/Flange material: 1, 3,4,E, G,K,M, or N.
j =Protection Class/dimension:
0 indicates painted aluminum enclosure,rated Type 4X and Type 6.
2 indicates stainless steel enclosure,rated Type 6P.
k=Cable: 2 (Type 6P), or any other alphanumeric digit(Enclosure types other than 6P).
1=Cable Length: (any alphanumeric digit).
m=Calibration: (any alphanumeric digit).
n=G-P Ring/Material: (any alphanumeric digit).
o=Special: 0.
p=Module/labeled: 1
Notes:
DQD 507 Rev.2009-09-01 Page:6
SR
CSA INTERNATIONAL
Certificate: 1665151 (LR 105802) Master Contract: 162767
Project: 2372993 Date Issued: November 24,2010
1) OPTIFLUX Primary Head must be used with the IFC 300 F-DIV2, IFC 300 F(/i)-DIV2, or IFC 090-F
DIV2 signal converters.
2) I.S.Electrodes when used with IFC 300 F-DIV2, IFC 300 F(/i)-DIV2, or IFC 090-F DIV2 signal
converters.
• OPTIFLUX 4000-DIV2 series Primary Head.Model VN034abcdefghijklmnop.
Operating ambient-50°C to 40°C; Temperature Code T6. Enclosure type as specified in the detail below.
Where:
a=Nominal Diameter: 1,2, 3,4, 5, 6, 7, 8,A,B, C, or D.
b=Nominal Pressure: 3,4, 5,A,B,D or E.
c=Approval: C.
d=Cable Connection: 1, 2,4,6,A,or C.
e=Converter Model: C, or D.
f=Lining: (any alphanumeric digit).
g=Electrode: 1,2, 3,4, 5, 6, 7,G,or U.
h=Construction of Electrodes: 1.
i=Housing/Flange material: 1,2, 3,4,A,B, C,D,K,L,M, or N.
j =Protection Class/dimension:
0 indicates painted aluminum enclosure,rated Type 4X and Type 6.
2 indicates stainless steel enclosure,rated Type 6P.
k=Cable: 2 (Type 6P), or any other alphanumeric digit(Enclosure types other than 6P).
1=Cable Length: (any alphanumeric digit).
in=Calibration: (any alphanumeric digit).
n=G-P Ring/Material: (any alphanumeric digit).
DQD 507 Rev.2009-09-01 Page:7
SR
CSA INTERNATIONAL
Certificate: 1665151 (LR 105802) Master Contract: 162767
Project: 2372993 Date Issued: November 24,2010
o=Special: 0.
p=Module/labeled: 1
Notes:
1) OPTIFLUX Primary Head must be used with the IFC 300 F-DIV2, IFC 300 F(/i)-DIV2, or IFC 090-F
DIV2 signal converters.
2) I.S.Electrodes when used with IFC 300 F-DIV2,IFC 300 F(/i)-DIV2,or IFC 090-F DIV2 signal
converters.
• OPTIFLUX 4000-DIV2 series Primary Head.Model VN044abcdefghijklmnop.
Operating ambient-50°C to 40°C;Temperature Code T6. Enclosure type as specified in the detail below.
Where:
a=Nominal Diameter: E,F, G,H,K, L,M, or N.
b=Nominal Pressure: 2, 3,4, 5,A,or B.
c=Approval: C.
d=Cable Connection: 1,2,4, 6, A,or C.
e=Converter Model: C, or D.
f=Lining: (any alphanumeric digit).
g=Electrode: 1, 3,4, 5, 6, 7, G,M,R, or U.
h=Construction of Electrodes: 1.
i=Housing/Flange material: 1, 3,4,A, C,D,K,M, or N.
j =Protection Class/dimension:
0 indicates painted aluminum enclosure,rated Type 4X and Type 6.
2 indicates stainless steel enclosure,rated Type 6P.
DQD 507 Rev.2009-09-01 Page:8
SR
CSA INTERNATIONAL
Certificate: 1665151 (LR 105802) Master Contract: 162767
Project: 2372993 Date Issued: November 24,2010
k=Cable: 2 (Type 6P), or any other alphanumeric digit(Enclosure types other than 6P).
1=Cable Length: (any alphanumeric digit).
m=Calibration: (any alphanumeric digit).
n=G-P Ring/Material: (any alphanumeric digit).
o=Special: 0.
p=Module/labeled: 1
Notes:
1) OPTIFLUX Primary Head must be used with the IFC 300 F-DIV2, IFC 300 F(/i)-DIV2, or IFC 090-F
DIV2 signal converters.
2) I.S.Electrodes when used with IFC 300 F-DIV2,IFC 300 F(/i)-DIV2,or IFC 090-F DIV2 signal
converters.
• OPTIFLUX 4000-DIV2 series Primary Head.Model VN054abcdefghijklmnop.
Operating ambient-50°C to 40°C;Temperature Code T6. Enclosure type as specified in the detail below.
Where:
a=Nominal Diameter: 28 inches,up to a maximum of 120 inches.
b=Nominal Pressure: 1,2, or A.
c=Approval: C.
d=Cable Connection: 1,2,4, 6,A, or C.
e=Converter Model: C, or D.
f=Lining: (any alphanumeric digit).
g=Electrode: 1, 3,4, 5, 6, 7, G,M, R, or U.
h=Construction of Electrodes: 1.
DQD 507 Rev.2009-09-01 Page:9
SR
CSA INTERNATIONAL
Certificate: 1665151 (LR 105802) Master Contract: 162767
Project: 2372993 Date Issued: November 24,2010
i=Housing/Flange material: 1, 3,4,A,C,D,K,M,or N.
j =Protection Class/dimension:
0 indicates painted aluminum enclosure,rated Type 4X and Type 6.
2 indicates stainless steel enclosure,rated Type 6P.
k=Cable: 2 (Type 6P), or any other alphanumeric digit(Enclosure types other than 6P).
1=Cable Length: (any alphanumeric digit).
in=Calibration: (any alphanumeric digit).
n=G-P Ring/Material: (any alphanumeric digit).
o=Special: 0.
p=Module/labeled: 1
Notes:
1) OPTIFLUX Primary Head must be used with the IFC 300 F-DIV2, IFC 300 F(/i)-DIV2, or IFC 090-F
DIV2 signal converters.
2) I.S.Electrodes when used with IFC 300 F-DIV2,IFC 300 F(/i)-DIV2,or IFC 090-F DIV2 signal
converters.
• OPTIFLUX 5000-DIV2 series Primary Head.Model VN184abcdefghijklmn.
Operating ambient-50°C to 40°C;Temperature Code T6.Enclosure type as specified in the detail below.
Where:
a=Nominal Diameter: 1,2, 3, 5, 6, 8,B, C, E, or F.
b=Nominal Pressure: 3,4, 5,A, or B.
c=Approval: C.
d=Cable Connection: 1,2,4, 6,A, or C.
DQD 507 Rev.2009-09-01 Page:10
SR
CSA INTERNATIONAL
Certificate: 1665151 (LR 105802) Master Contract: 162767
Project: 2372993 Date Issued: November 24,2010
e=Converter Model: C, or D.
f=Mounting Material: (any alphanumeric digit).
g=G-Ring Material/Gasket Material: (any alphanumeric digit).
h=Electrode: 0.
i=Protection Class/dimension:
0 indicates painted aluminum enclosure,rated Type 4X and Type 6.
2 indicates stainless steel enclosure,rated Type 6P.
j =Cable: 2(Type 6P),or any other alphanumeric digit(Enclosure types other than 6P).
k=Cable Length: (any alphanumeric digit).
1=Calibration: (any alphanumeric digit).
in=Special: 0.
n=Module/labeled: 1
Notes:
1) OPTIFLUX Primary Head must be used with the IFC 300 F-DIV2,IFC 300 F(/i)-DIV2, or IFC 090-F
DIV2 signal converters.
2) I.S.Electrodes when used with IFC 300 F-DIV2,IFC 300 F(/i)-DIV2, or IFC 090-F DIV2 signal
converters.
• OPTIFLUX 5000-DIV2 VN244abcdefghijklmnop Primary Head. (Flange Version) Operating ambient-50°C
to 40°C; Temperature Code T6. Enclosure type as specified in the detail below.
Where:
a=Nominal diameter: 2,4,6,7,A,B.
b=Nominal pressure: 3, 5 or A.
c=Approval: C.
DQD 507 Rev.2009-09-01 Page:11
SP
CSA INTERNATIONAL
----------------------
Certificate: 1665151 (LR 105802) Master Contract: 162767
Project: 2372993 Date Issued: November 24,2010
d=Cable connection: 1, 2,4, 6,A or C.
e=Converter model: C or D.
f =Lining: not safety relevant.
g=Electrode: 0.
h=Construction of electrodes: 1.
i =Housing/flange material: 0
j =Protection class/dimension: 0 or 2.
k=Cable: 2 (Type 6P); Other—Not Safety Relevant.
1 =Cable length: not safety relevant.
m=Calibration: not safety relevant.
n=G-P-Ring/Material: not safety relevant.
o=Special: 0.
p=Module/labeled: 0 or 1.
Notes:
1) OPTIFLUX Primary Head must be used with the IFC 300 F-DIV2, IFC 300 F(/i)-DIV2, or IFC 090-F
DIV2 signal converters.
2) I.S. Electrodes when used with IFC 300 F-DIV2, IFC 300 F(/i)-DIV2, or IFC 090-F DIV2 signal
converters.
• OPTIFLUX 6000-DIV2 series Primary Head.Model VN224abcdefghijklmn.
Operating ambient-50°C to 40°C; Temperature Code T6. Enclosure type as specified in the detail below.
Where:
a=Nominal Diameter: 1,2, 3, 5, or 6.
DQD 507 Rev.2009-09-01 Page:12
SR
CSA INTERNATIONAL
Certificate: 1665151 (LR 105802) Master Contract: 162767
Project: 2372993 Date Issued: November 24,2010
b=Connection: 5, 6,A,B,H,K,L,M,N,P, S, T,or U.
c=Approval: C.
d=Cable Connection: 1,2,4, 6,A, or C.
e=Converter Model: C, or D.
f=Mounting Material: (any alphanumeric digit).
g=G-Ring Material/Gasket Material: (any alphanumeric digit).
h=Electrodes: 1, 3,4, 5, 6, 7,or A.
i=Protection Class/dimension:
0 indicates painted aluminum enclosure,rated Type 4X and Type 6.
2 indicates stainless steel enclosure,rated Type 6P.
j =Cable: 2(Type 6P),or any other alphanumeric digit(Enclosure types other than 6P).
k=Cable Length: (any alphanumeric digit).
1=Calibration: (any alphanumeric digit).
m=Special: 0.
n=Module/labeled: 1
Notes:
3) OPTIFLUX Primary Head must be used with the IFC 300 F-DIV2,IFC 300 F(/i)-DIV2, or IFC 090-F
DIV2 signal converters.
4) I.S.Electrodes when used with IFC 300 F-DIV2, IFC 300 F(/i)-DIV2, or IFC 090-F DIV2 signal
converters.
• Primary Head Model IFS 4000 F-DIV2 series,magnetic flow meter sensor,rated 40 Vdc, 125mA. Install per
Dwgs 8.30702.63 and 8.30702.64.Temperature Code T6,Ta=-50°C to 40°C.
Notes:
DQD 507 Rev.2009-09-01 Page:13
SP
CSA INTERNATIONAL
----------------------
Certificate: 1665151 (LR 105802) Master Contract: 162767
Project: 2372993 Date Issued: November 24,2010
1) Primary Head,Model IFS 4000 F-DIV2 must only be used with Certified remote signal converter models
IFC 300 F-DIV2,IFC 300 F(/i)- DIV2 or IFC 090-F DIV2.
2) I.S. Electrodes when used with remote signal converters,models IFC 300 F-DIV2,IFC 300 F(/i)-DIV2, or
IFC 090-F DIV2
3) Internal order code known as type V303.
• Primary Head Model IFS 5000 F-DIV2 series,magnetic flow meter sensor,rated 40 Vdc, 125mA. Install per
Dwgs 8.30702.63 and 8.30702.64. Temperature Code T6,Ta=-50°C to 40°C.
Notes:
1) Primary Head Model IFS 5000 F-DIV2 must only be used with Certified remote signal converter, models
IFC 300 F-DIV2,IFC 300 F(/i)-DIV2 or IFC 090.
2) I.S. Electrodes when used with remote signal converters,models IFC 300 F-DIV2,IFC 300 F(/i)-DIV2, or
IFC 090-F DIV2
3) Internal order code known as type V304.
• IFC300 series Signal Converter. Model a VN304bcdefghijklmno. Rated 100-230Vac,22VA, 50-60 Hz and/or
12-24Vdc/24V ac/dc, 12W. Operating ambient-50°C to 40°C; Temperature Code T6. Enclosure type 4X,Type
6.
Where:
a=Input/Output:
F(/i)-DIV2 indicates associated nonincendive field wiring and I.S. entity parameters,per control drawing
8.30702.65.
F-DIV2 indicates normally nonarcing field wiring for use with Class I Div 2 wiring methods,
b=Type: 4, or H.
c=Power Supply: 1,4, or A.
DQD 507 Rev.2009-09-01 Page:14
SR
CSA INTERNATIONAL
Certificate: 1665151 (LR 105802) Master Contract: 162767
Project: 2372993 Date Issued: November 24,2010
d=Approval: C.
e=Cable Connection: 4,or 6.
f=Instruction/Programming Language: (any alphanumeric digit).
g=Custody Transfer: 0.
h=Process Diagnostics: (any alphanumeric digit).
i=Converter Housing:
1 indicates painted aluminum enclosure,rated Type 4X and Type 6.
2 indicates stainless steel enclosure,rated Type 4X and Type 6.
j =Labeled: 1.
k=Communications Slot I01 Selection: 1,2, 3,4, 6,7, 8,B, C,D,E,F, G, or H.
1=First IO Module(102/A): 0, 1,2, 8,A,B, C,E,F, G,H, or K.
m=Second IO Module(IO2/13): 0, 8,A,B, C,E,F, G,H, or K.
n=Reference Method: 0,or 1
o=Tagplate: (any alphanumeric digit).
Notes:
1) OPTIFLUX IFC300 F signal converter electronics must be installed in explosionproof MH300C housing,
and used with OPTIFLUX,IFS 4000E-DIV2 or IFS 5000E-DIV2 Primary Head.
2) Install per manufacturers drawings 8.30702.63 or 8.30702.64.
• OPTIFLUX a300 C-DIV2 Flowmeter. Rated 100-230Vac,22VA, 50-60 Hz and/or 12-24Vdc/24V ac/dc,
12W. Operating ambient-50°C to 40°C; Temperature Code T6. Enclosure type 4X, Type 6.
Where:
a=Type: 1,2, 4, 5, or 6.
DQD 507 Rev.2009-09-01 Page:15
SP
CSA INTERNATIONAL
----------------------
Certificate: 1665151 (LR 105802) Master Contract: 162767
Project: 2372993 Date Issued: November 24,2010
Consists of the OPTIFLUX a000-DIV2 series Primary Head(Where: a=Type: 1, 2, 4, 5, or 6), and the IFC300
F-DIV2 VN304bcdefghijklmno Signal Converter.
• OPTIFLUX a300 C/i-DIV2 Flowmeter. Rated 100-230Vac, 22VA, 50-60 Hz and/or 12-24Vdc/24V ac/dc,
12W. Operating ambient-50°C to 40°C; Temperature Code T6. Enclosure type 4X, Type 6.
Where:
a=Type: 1,2,4, 5, or 6.
Consists of the OPTIFLUX a000-DIV2 series Primary Head(Where: a=Type: 1,2,4, 5, or 6), and the IFC300
F(/i)-DIV2 VN304bcdefghijklmno Signal Converter.
Note:
1) Nonincendive Field Wiring and Intrinsically Safe Entity Parameters when installed per manufacturers
Control Drawings and drawing 8.30702.65.
APPLICABLE REQUIREMENTS
CAN/CSA Standard C22.2 No. 0-M91 (Reaffirmed 2001)-General Requirements-Canadian Electrical Code,Part II
CAN/CSA Standard C22.2 No. 94-M94(Reaffirmed 2001)-Special Purpose Enclosures
CSA Standard C22.2 No. 142-M1987(Reaffirmed 2000)-Process Control Equipment
CSA Standard C22.2 No.25-1966(Reaffirmed 2000)-Enclosures for Use in Class II Groups E,F,and G Hazardous
Locations
CSA Standard C22.2 No.30-M1986(Reaffirmed 2003)-Explosion-Proof Enclosures for Use in Class I Hazardous
Locations
CAN/CSA Standard C22.2 No. 157-92(Including update No.2,June,2003)-Intrinsically Safe and Non-Incendive
Equipment for Use in Hazardous Locations.
CSA Standard C22.2 No.213-M1987(Reaffirmed 2008)-Non-incendive Electrical Equipment for Use in Class I,Division
2 Hazardous Locations
DQD 507 Rev.2009-09-01 Page:16