HomeMy WebLinkAboutAP_WATERFLUX 3070__en_MCERTS Water Monitoring UK___ �01A�
q$, Environment
s ra
a f � � MC E R Agency
PRODUCI CVRTIFICATION THE ENVIRONMENT AGENCY'S
Oil MONITORING CERTIFICATION SCHEME
PRODUCT CONFORMITY CERTIFICATE
This is to certify that the
WA TERFLUX 3070
manufactured by:
KROHNE Altometer
A production facility of KROHNE AG, Basel
Kerkeplaat 12
3313 LC Dordrecht
The Netherlands
has been assessed by Sira Certification Service
and for the conditions stated on this certificate complies with:
MCERTS Performance Standards for Water Monitoring
Equipment Part 3, Version 2.1 dated March 2008
Certification Ranges
Size range DN25 to DN300
Project No: 16W22202
Certificate No: Sira MC100178/00
Initial Certification: 08 November 2010
This Certificate Issued 08 November 2010
Renewal Date: 07 November 2015
Technical Director
MCERTS is operated on behalf of the Environment Agency by
Sira Certification Service
12 Acorn Industrial Park, Crayford Road, Crayford
Dartford, Kent, UK, DA1 4AL
Tel: 01322 520500 Fax: 01322 520501
This certificate may only be reproduced in its entirety and without change
To authenticate the validity of this certificate please see www.siraenvironmental.com
Page 1 of 6
q$,
Environment
s raUKAS MC E RTS Agency
rofication Service
PRODUCI
CVRTIFICATION THE ENVIRONMENT AGENCY'S
OilI MONITORING CERTIFICATION SCHEME
Approved Site Application
The product may be used on all MCERTS applications including abstraction, effluent discharge,
ultraviolet disinfection and industrial processing.
Any potential user should ensure, in consultation with the manufacturer, that the monitoring system is
suitable for the intended application. For general guidance on monitoring techniques refer to the
Environment Agency Monitoring Technical Guidance Notes available at www.mcerts.net.
Basis of Certification
This certification is based on the following Test Report(s) and on Sira's assessment and ongoing
surveillance of the product and the manufacturing process:
Sira Report 16W22202 dated 08 November 2010
NMi Certain B.V Report R49-1 /2006-NL1-09.01; 812577, dated March 2009
NMi Certain B.V Report R49-1 /2006-NL1-10.01; 9200759, dated November 2009
Product Certified
The measuring system consists of the following parts: Pipe Flow rate Unit
size max min
• WATERFLUX 3000 electromagnetic flow / water DN25 16 0.32 m3/hr
meter DN40 25 0.50 m3/hr
• IFC 070 electromagnetic signal converter DN50 40 0.81 m3/hr
(C/compact or F/ remote version) DN65 63 1.30 m3/hr
DN80 100 2.00 m3/hr
This certificate applies to all instruments fitted with software DN100 160 3.20 m3/hr
version 4.0.4. (serial number Al 01 xxxxx onwards). DN125 250 5.00 m /hr
DN150 400 8.10 m3/hr
DN200 630 13.00 m3/hr
DN250 1000 20.00 m3/hr
DN300 1600 32.00 m3/hr
Certificate No: Sira MC100178/00
This Certificate Issued: 08 November 2010
This certificate may only be reproduced in its entirety and without change
To authenticate the validity of this certificate please see www.siraenvironmental.com
Page 2 of 6
q$,
Environment
s raUKAS MC E RTS Agency
rofication Service
PRODUCI
CVRTIFICATION THE ENVIRONMENT AGENCY'S
Oil MONITORING CERTIFICATION SCHEME
Certified Performance
The instrument was evaluated for use under the following conditions:
Ambient Temperature Range: -100C to+55 °C
The instrument meets MCERTS Class 1 requirements for the combined performance characteristic as specified in
Table 6 of the MCERTS performance standard. Details of individual performance characteristics are summarised
below:
Results are expressed as error%reading, unless otherwise stated.
Test Results expressed as error% Other results MCERTS
of reading specification
<0.5 <1.0 <1.5 <2.0
Protection against unauthorised access A custody transfer seal is present Clause 3.1.2
Indicating device The flowmeter incorporates an indicating device, Clause 3.1.3
analogue and digital output signal
Units of measurement Various units of measurement are available and Clause 3.1.6
displayed.
Bi-directional flow The flowmeter displays a'+'or'—'flow reading Clause 3.1.8
Clause 6.3.2
Combined performance characteristic 0.70 ±1.5%Class 1
Mean error
DN25 0.46
DN50 0.44
DN80 0.29 Clause 6.3.2
DN100 0.50 ±1.5%Class 1
DN200 0.47
DN300 0.33
Mean error at low flow
DN25(0.025m3/h) 1.33
DN50(0.11m3/h) 1.43 Supplementary
DN100(0.62m3/h) 1.35 data
DN200(4.01 m3/h) 0.58
DN300(3.97m3/h) 2.83
Clause 6.3.2
Repeatability 0.15 1%Class 1
Clause 6.3.3
Supply voltage <0.02 2.9V to 3.6V 0.5%Class 1
Certificate No: Sira MC100178/00
This Certificate Issued: 08 November 2010
This certificate may only be reproduced in its entirety and without change
To authenticate the validity of this certificate please see www.siraenvironmental.com
Page 3 of 6
q$,
Environment
s raUKAS MC E RTS Agency
rofication Service
PRODUCI
CVRTIFICATION THE ENVIRONMENT AGENCY'S
Oil MONITORING CERTIFICATION SCHEME
Test Results expressed as error% Other results MCERTS
of reading specification
<0.5 <1.0 <1.5 <2.0
Clause 6.3.5
Fluid Temperature 0.22 12°C to 500C 0.5%Class 1
Clause 6.3.6
Ambient air temperature 0.04 -10°C to+55°C 0.5%Class 1
Test conducted Clause 6.3.6
Relative humidity -0.03 at 45%relative 0.5%Class 1
humidity
Clause 6.3.9
Stray currents <0.02 0.5%Class 1
Bi-directional flow Mean error
Mean error -0.08
±1.5%Class 1
Repeatability See Note 1 Repeatability
1/o Class 1
No changes in pre set data Clause 6.3.1
Loss of Power for electronic flowmeters to be reported
Clause 6.3.19
Response time <30s 30 seconds
Clause 6.1.2
Warm up time <1s to be reported
Clause 6.3.20
Vibration -0.11 Note 2 to be reported
Note 1: Repeatability for bi-directional flow could not be calculated as only 2 data points were taken at each flow rate
Note 2: Vibration test conducted with random vibrations over frequency range 10Hz to 150Hz on 3 perpendicular axes, for a
period of at least 2 minutes per axis.
Note 3: The following tests are not applicable to the flowmeter:
6.3.4 Output Impedance 6.3.14 Flow reversal
6.3.7 Incident light 6.3.15 Ancillary devices
6.3.8 Sensor location 6.3.16 Effect of conduit material
6.3.10 Sonic velocity compensation&response 6.3.17 Effect of conduit size
6.3.11 Accuracy of computation 6.3.18 Fill level
6.3.12 User defined stage-discharge equation
Certificate No: Sira MC100178/00
This Certificate Issued: 08 November 2010
This certificate may only be reproduced in its entirety and without change
To authenticate the validity of this certificate please see www.siraenvironmental.com
Page 4 of 6
q$,
Environment
s raUKAS MC E RTS Agency
rofication Service
PRODUCI
CVRTIFICATION THE ENVIRONMENT AGENCY'S
OilI MONITORING CERTIFICATION SCHEME
Field Test Results
The field test was conducted on a WATERFLUX 3070 in series with a mechanical water meter for 3
months at a water utility site.
Test Results expressed as error% Other results MCERTS
of reading specification
<0.5 <1.0 <1.5 <2.0
Error range 1.40%to 7.52% Clause 7.3
Error under field test conditions
Field test error is<2%for 42%of readings 2%Class 1
Field test error is<5%for 88%of readings 5%Class 2
Field test error is<8%for 100%of readings 8%Class 3
Up time 100% Clause 7.4
>95%
Maintenance none Clause 7.5
to be reported
Certificate No: Sira MC100178/00
This Certificate Issued: 08 November 2010
This certificate may only be reproduced in its entirety and without change
To authenticate the validity of this certificate please see www.siraenvironmental.com
Page 5 of 6
q$,
Environment
s raUKAS MC E RTS Agency
rofication Service
PRODUCI
CVRTIFICATION THE ENVIRONMENT AGENCY'S
Oil MONITORING CERTIFICATION SCHEME
Description:
The WATERFLUX 3070 is a battery powered water meter based on the electromagnetic flow principle.
It is dedicated for applications in the water and wastewater industry. It consists of a WATERFLUX 3000
sensor and a battery operated IFC 070 converter. The converter can be mounted directly on the sensor
(compact version)or separately(field version).
WATERFLUX 3070 is designed for custody transfer applications. It meets the requirements of the OIML
R49 and can be verified according to MI-001. For potable water applications it is certified with DVGW,
ACS, KTW. Krohne Altometer meets the applicable requirements of MID module D for the conformity
assessment of water meters. The accuracy of the WATERFLUX 3070 is 0.2 % of the measured value
plus 0.5 mm/s and every flow meter leaving the factory is calibrated.
The principle of the electromagnetic flow metering is based on Faraday's law of induction: passing an
electrically conductive body through a magnetic field, a voltage is induced. This voltage is proportional
to velocity and picked up by electrodes.
The WATERFLUX 3000 sensor has a rectangular cross section resulting in an improved flow profile,
high accuracy in low flow conditions, large span and minimal power consumption. It is stated by the
manufacturer that straight up-and downstream lengths are not required.
A grounding electrode provides the grounding of the medium, grounding rings are not required. The
field version of the sensor can be submerged (IP68) and with an optional buriable coating can be
installed directly underground. The tube and coil are designed for efficient energy consumption.
The IFC 070 converter is battery operated with two internal lithium batteries, with an operating life time
is up to 15 years. It can display positive and negative counter, sum counter and flow rate. It can further
provide diagnostic information for self checking, battery warnings, counter overrun, flow direction and
empty pipe detection. It has two pulse outputs and two status outputs.
General Notes
1. This certificate is based upon the equipment tested. The Manufacturer is responsible for
ensuring that on-going production complies with the standard(s) and performance criteria defined
in this Certificate. The Manufacturer is required to maintain an approved quality management
system controlling the manufacture of the certified product. Both the product and the quality
management system shall be subject to regular surveillance according to 'Regulations Applicable
to the Holders of Sira Certificates'. The design of the product certified is defined in the Sira
Design Schedule for certificate No. Sira MC100178/00
2. If certified product is found not to comply, Sira Certification Service should be notified
immediately at the address shown on this certificate.
3. The Certification Marks that can be applied to the product or used in publicity material are
defined in `Regulations Applicable to the Holders of Sira Certificates'.
4. This document remains the property of Sira and shall be returned when requested by the
company.
Certificate No: Sira MC100178/00
This Certificate Issued: 08 November 2010
This certificate may only be reproduced in its entirety and without change
To authenticate the validity of this certificate please see www.siraenvironmental.com
Page 6 of 6