HomeMy WebLinkAboutAP_WATERFLUX 3070__en_MI-001_100409__ M � EC - type examination
certificate
Number T10201 revision 0
Project number: 812577
Page 1 of 1
Issued by NMi Certin B.V.
Hugo de Grootplein 1
3314 EG Dordrecht
The Netherlands
In accordance with The Council Directive 2004/22/EC on Measuring Instruments (MID).
Applicant KROHNE Altometer
Kerkeplaat 12
3313 LC Dordrecht
The Netherlands
In respect of A model of a water meter
Manufacturer KROHNE Altometer
Type WATERFLUX
Characteristics Measuring range See paragraph 1.2 of the description.
Mechanical environmental class M2
Electromagnetic environment class E2
Temperature range liquid +0,1 °C/+50 °C
Temperature range ambient -10 °C/+55 °C
In the description T10201 revision 0 further characteristics are described.
Valid until 10 September 2019
Description and The water meter is described in the description number T10201 revision 0 and
Documentation documented in the documentation folder number. T10201-1, appertaining to
this EC-type examination certificate.
)N � d Body no. 0122
, 10 September 2009
an
Head Certification Board
NMi Certin BV Parties concerned can lodge objection This document is issued under the provision
Hugo de Grootplein 1 against this decision,within six weeks after that no liability is accepted and that the
3314 EG Dordrecht the date of submission,to the general applicant shall indemnify third-party ' \�
PO Box 394 manager NMi(see"Regulation objection liability. 'N
and appealal against decisions of NMi") \``v I
3300 AJ Dordrecht,NL Reproduction of the complete document is
T+31 78 6332332 permitted.
F+31 78 6332309 I NSKETIO N
certin®nmi.nl vvww.nmi.nl RvA 1 172
Mil
Description
Number T10201 Revision 0
Project number 812577
Page 1 of 5
1 General information on the water meter
All properties of this water meter, whether mentioned or not, may not be in conflict with
the legislation. The measuring part of the electromagnetic water meter consists of a
cylindrical piece of piping with a square-like flow profile and 2 magnetic spools and 3
electrodes and a Signal Converter, which is used to drive the magnetic spools and convert
the electrode voltage to a flow rate.
1.1 Essential Parts
Description Document Remarks
Measurement 10201-01 The inside of the cylindrical measuring tube is
sensor provided with a square-like flow profile. The
internal flow sensor is covered with an
insulating liner. Through the liner the 3
electrodes are in contact with the liquid. The
flow sensor is equipped with 2 magnetic coils.
printed circuit 10201-05 IFC 070 flow converter
board 10201-06 Signal Converter and meter readin
Mi
Description
Number T10201 Revision 0
Project number 812577
Page 2 of 5
1.2 Essential Characteristics
1.2.2 Flow characteristics
Meter Flow rates [m3/h]
size Minimum Transitional Permanent Overload
Q1 Q2 Q3 Q4
DN50 01158 0,252 63 78,75
0,10 0,16 40 50
DN65 0,25 0,40 100 125
0,16 0,25 63 78,8
0,40 0,64 160 200
DN80
0,25 0,40 100 125
DN100 0,625 1,00 250 312,5
0,40 0,64 160 200
1,00 1,60 400 500
DN125
0,63 1,00 250 312,5
1,575 2,52 630 787,5
DN150
1,00 1,60 400 500
DN200 2,54 1 4,06 800 1000
1,58 2,52 630 787,5
1.2.3 Measuring principle
In documentation 10201-01, paragraph 8.1, the measuring principle is described.
1.2.4 Bi-directional flow
The sensor can be used to measure flow in forward and reverse directions. See also the
conditions as stated in chapter 3.
1.2.5 Software specification (refer to WELMEC guide 7.2):
• Software type P;
• Risk Class C;
• Extension L, while extensions D, S and T are not applicable.
Software version Remarks
4.0.4 The software version is
displayed after operating
the right infrared key.
Mi
Description
Number T10201 Revision 0
Project number 812577
Page 3 of 5
1.2.6 Operation and presentation of legal data
The meter is standard equipped with an electronic display and can be operated using the
infrared keys on the display module. The operation is further described in Chapter 6 of
the Manual 10201-01.
The measured volume is presented by means of an electronic display
The display register is built up as follows:
Meter Indicating range Verification scale
Size (minimum value) interval
[m3] (maximum value)
[mi]
DN50 99.999 0,0001
DN80 until DN150 999.999 0,001
DN200 9.999.999 0,001
1.2.7 Accountable alarms
If malfunctions are detected a visible alarm will be performed which remains present
until the alarm is acknowledged and the cause of the alarm is suppressed.
During the accountable alarms the registration of the volume in the totalizers is stopped.
The validity of the program and the parameters is continuously checked. If these checks
fail an alarm is generated.
1.2.8 Operating pressure range
The measuring tube and measuring sensor may be used at an absolute pressure up to 10
bar.
1.3 Essential Shapes
1.3.2 Markings:
- The markings have to fulfil the requirements stated in the legislation.
- For the remote version:
■ The serial number of the signal converter is mentioned on the measurement
sensor and/or the serial number of the measurement sensor is mentioned on the
signal converter.
An example of the markings is given in document no. 10201-04.
M1
Description
Number T10201 Revision 0
Project number 812577
Page 4 of 5
1.4 Conditional parts
1.4.1 Housing
The housing of the Signal Converter consists of a metallic housing (compact and remote
version).
1.4.2 Cabling remote reading
When the meter is executed as a remote version the cable between the measuring sensor
and reading unit should be mounted fixed. During use it should not be possible to move
this cable.
1.4.3 Power supply
The water meter is powered by one or two 3,6 Volts D-cell Lithium primary (non-
rechargeable) batteries. The replacement date of the battery shall be indicated on the
meter.
A low battery alarm is generated one year before the meter stops functioning (based on
actual settings and measuring situation).
1.5 Conditional characteristics
1.5.1 Programming
Essential parameters (see documentation 10201-01) cannot be changed after the access
control function (menu item 13) is set for "fiscal metering". In order to disable the "fiscal
metering" a seal has to be broken, see documentation 10201-01 page 19 and 20.
1.5.2 Low flow cut off
The low flow cut off is a programmable minimum flow.
If the flow measured by the meter is below the programmed low flow cut off, the flow
will be considered zero, so the meter reading will no longer change
The programmed value cannot not exceed 10 mm/s, see documentation no. 10201-01
page 20.
1.5.3 Cylindrical pipe with a square-like flow profile
The cylindrical pipe and the internal square flow profile is constructed so that the
combination of material and wall thickness is such that the influence due to changing
liquid pressure is negligible.
1.6 Non essential Shape
1.6.1 Alarm or pulse outputs
MI
Description
Number T10201 Revision 0
Project number 812577
Page 5 of 5
2 Seals
The following seals are applied:
- The data plate is fixed to the water meter and secured against removal by sealing or
will be destroyed when removed.
- The measurement sensor against opening.
- The cover of the display of the signal converter against opening. This makes it
impossible to "unlock" the access control function (menu item 13). For unlocking this
function, it is necessary to activate a hardware switch which is located inside the
electronics unit. See documentation no. 10201-01 page 19.
See the drawing no. 10201-03 for an example of the sealing.
The securing component has to bear either:
— A mark of the manufacturer laid down in an approved quality system by a Notified
body, or;
— A mark of a Notified body.
3 Conditions for conformity assessment
3.1 Bi-directional flow measurement
During conformity assessment it is sufficient to verify a bi-directional meter only in one
direction.
M � EC type - examination
certificate
Number T10201 revision 1
Project number: 9200759
Page 1 of 1
Issued by NMi Certin B.V.,
designated and notified by the Netherlands to perform tasks with respect to
conformity modules mentioned in article 9 of Directive 2004/22/EC, after
having established that the Measuring instrument meets the applicable
requirements of Directive 2004/22/EC, to:
Manufacturer KROHNE Altometer
Kerkeplaat 12
3313 LC Dordrecht
The Netherlands
Measuring instrument A water meter
Type Waterflux
Destined for the measurement of Water
Measuring Range See paragraph 1.2 of the
description.
Environment classes M2/E2
Temperature range liquid +0,1 °C/+50 °C
Temperature range ambient -10 °C/+55 °C
Further properties are described in:
— Description T10201 revision 1;
— Documentation folder T10201-1.
Valid until 10 September 2019
Remarks This revision 1 replaces the previous version 0. The documentation folder
was not changed.
Issuing Authority NMi qqrtin B.V., Notified Body number 0122
9 ApVO.
Head of the Certification Board
NMi Certin B.V. This document is issued under the provision that Parties concerned can lodge '
Hugo d �/ `��
e Grootplein 1 no liability is accepted and that the applicant objection against this decision,within �c Y
3314 EG Dordr'echt shall indemnify third-party liability. six weeks after the date of `A\
The Netherlands submission,to the general manager I` 7
T+31 78 6332332 The designation of NMi Certin BV.as Notified of NMi(see www.nmi.nl). INSPECT(DN
certin@nmi.nl Body can be verified at http:// q
www.nmi.nl ec.europa.eu/enterprise/newapproach/nando/ Reproduction of the complete RuA ( 12L
document only is permitted.
MI Description
Number T10201 Revision 1
Project number 9200759
Page 1 of 4
1 General information on the water meter
All properties of this water meter, whether mentioned or not, shall not be in conflict
with the legislation. The measuring part of the electromagnetic water meter consists of a
cylindrical piece of piping with a square-like flow profile and 2 magnetic spools and
3 electrodes and a Signal Converter, which is used to drive the magnetic spools and
convert the electrode voltage to a flow rate.
This revision 1 is issued due to:
- Addition of DN25, DN40, DN250 and DN300 sizes.
- Some editorial changes.
1.1 Essential Parts
Description Document Remarks
Measurement 10201-01 The inside of the cylindrical measuring tube is
sensor provided with a square-like flow profile. The
internal flow sensor is covered with an
insulating liner. Through the liner, the
3 electrodes are in contact with the liquid. The
flow sensor is equipped with 2 magnetic coils.
printed circuit 10201-05 IFC 070 flow converter
board 10201-06 Signal Converter and meter reading
1.2 Essential Characteristics
1.2.1 Flow characteristics
Flow rates [m3/h]
Meter
size Minimum Transitional Permanent Overload
Q1 Q2 Q3 Q4
0,04 0,064 16 20
DN25
0,025 0,04 10 12,5
0,10 0,16 40 50
DN40
0,063 0,1 25 31,3
0,158 0,252 63 78,75
DN50
0,10 0,16 40 50
0,25 0,40 100 125
DN65
0,16 0,25 63 78,8
0,40 0,64 160 200
DN80
0,25 0,40 100 125
DN100 0,625 1,00 250 312,5
0,40 0,64 160 200
MI Description
Number T10201 Revision 1
Project number 9200759
Page 2 of 4
Flow rates [m3/h]
Meter
size Minimum Transitional Permanent Overload
Q1 Q2 Q3 Q4
1,00 1,60 400 500
DN125
0,63 1,00 250 312,5
1,575 2,52 630 787,5
DN 150
1,00 1,60 400 500
2,54 4,06 800 1000
DN200
1,58 2,52 630 787,5
DN250 2,5 4 1000 1250
DN300 4,0 6,4 1600 2000
1.2.2 Measuring principle
In documentation 10201-01, paragraph 8.1, the measuring principle is described.
1.2.3 Bi-directional flow
The sensor can be used to measure flow in forward and reverse directions. See also the
conditions as stated in chapter 3.
1.2.4 Software specification (refer to WELMEC guide 7.2):
• Software type P;
• Risk Class C;
• Extension L, while extensions D, S and T are not applicable.
Software versions Remarks
4.0.4 The software version is displayed after
operating the right-hand infrared key.
1.2.5 Operation and presentation of legal data
The meter is standard equipped with an electronic display and can be operated using the
infrared keys on the display module. The operation is further described in Chapter 6 of
the Manual 10201-01.
The measured volume is presented by means of an electronic display
The display register is built up as follows:
Indicating range Verification scale interval
Meter (minimum value) (maximum value)
Size [mi] [m3]
DN25, DN40 and DN50 99.999 0,0001
DN65, DN80, DN100, 999.999 0,001
DN125 and DN150
DN200, DN250 and 9.999.999 0,01
DN300
M � Description
Number T10201 Revision 1
Project number 9200759
Page 3 of 4
1.2.6 Accountable alarms
If malfunctions are detected a visible alarm will be displayed which remains present until
the alarm is acknowledged and the cause of the alarm is suppressed.
During the accountable alarms the registration of the volume in the totalizers is stopped.
The validity of the program and the parameters is continuously checked. If these checks
fail, an alarm is generated.
1.2.7 Maximum operating pressure:
- 16 bar(g) for sizes DN150 and smaller;
- 10 bar(g) for sizes DN200 and larger.
1.3 Essential Shapes
1.3.1 Markings:
- The markings have to fulfil the requirements stated in the legislation.
- For the remote version:
• The serial number of the signal converter is mentioned on the measurement sensor
and/or the serial number of the measurement sensor is mentioned on the signal
converter.
An example of the markings is given in document no. 10201-04.
1.4 Conditional parts
1.4.1 Housing
The housing of the Signal Converter consists of a metallic housing (compact and remote
version).
1.4.2 Cabling remote reading
When the meter is executed as a remote version the cable between the measuring sensor
and reading unit should be mounted fixed. During use it should not be possible to move
this cable.
1.4.3 Power supply
The water meter is powered by one or two 3,6 Volts D-cell Lithium primary (non-
rechargeable) batteries. The replacement date of the battery shall be indicated on the
meter.
A low battery alarm is generated one year before the meter stops functioning (based on
actual settings and measuring situation).
1.5 Conditional characteristics
1.5.1 Programming
Essential parameters (see documentation 10201-01) cannot be changed after the access
control function (menu item 13) is set for "fiscal metering". In order to disable the "fiscal
metering" a seal has to be broken, see documentation 10201-01 page 19 and 20.
1.5.2 Low flow cut off
The low flow cut off is a programmable minimum flow. If the flow measured by the
meter is below the programmed low flow cut off, the flow will be considered zero and
the meter reading will no longer change
The programmed value cannot not exceed 10 mm/s, see documentation no. 10201-01
page 20.
MI Description
Number T10201 Revision 1
Project number 9200759
Page 4 of 4
1.5.3 Cylindrical pipe with a square-like flow profile
The cylindrical pipe and the internal square flow profile is constructed so that the
combination of material and wall thickness is such that the influence due to changing
liquid pressure is negligible.
1.6 Non essential Shape
- Alarm or pulse outputs
2 Seals
The following seals are applied:
- The data plate is fixed to the water meter and secured against removal by sealing or
will be destroyed when removed.
- The measurement sensor against opening.
- The cover of the display of the signal converter against opening. This makes it
impossible to "unlock" the access control function (menu item 13). For unlocking this
function, it is necessary to activate a hardware switch which is located inside the
electronics unit. See documentation no. 10201-01 page 19.
See the drawing no. 10201-03 for an example of the sealing.
The securing component has to bear either:
- A mark of the manufacturer laid down in an approved quality system by a Notified
body, or;
- A mark of a Notified body.
3 Conditions for conformity assessment
- Bi-directional flow measurement
During conformity assessment it is sufficient to verify a bi-directional meter only in
one direction.